top of page

3 Hidden Causes of Implant Prosthetic Complications

Restoring dental implants is not easy.

There are a whole host of headaches awaiting us if we don’t follow certain rules.  Prosthetic implant complications can be neatly filed into esthetic, hygienic, and functional categories.  In this post we’ll review the various functional complications and the three rules that will keep us out of trouble.


Normal occlusal forces should not cause any of the above functional complications.  So what causes the harmful stress forces?  There are three possible sources that may be hard to detect.  Let’s sort them out.


Normal occlusal forces should be sustained by our implant solution.  However the patient may clench, brux, or have a silly parafunctional habit that introduces abnormal, stressful forces into the system.  I try to detect these clues during the treatment planning phase.  But perhaps the best way to find out how our patients will function on our restorations is to have them wear an implant-supported provisional before the final is made.  I’ve been doing this more and more in my practice since it will also help me evaluate potential esthetic and hygienic complications.  If I’m concerned a patient will overload the restoration, I will reduce the occlusion on the final crown to help protect it.  For a full discussion on implant protective occlusion, check out this post.

Note the open margins on the mesial aspects of the abutments. This framework also failed the quarter-turn test for passive fit.


(2) Misfit

This is a big problem.  Abutments must be fully and passively seated into the implant.  If not, even normal occlusal loads may not be transferred down the long axis of the implant, but rather to a weak link in the system.  That weak link could be the abutment screw, prosthetic screw, cement, etc.  That’s in the best case scenario.  The worst case scenario is that the normal occlusal forces are transferred to the implant and cause excessive peri-implant bone loss, implant fracture, or other nightmares.  Check for passive fit and make sure you have complete seating.

(3) Poor Design

Normal occlusal forces can overwhelm our implant solution if our solution is under-engineered.  Small diameter implants, short implants, and large cantilevers are some examples of engineering decisions that may end up failing us if used in the wrong situation. Proper treatment planning and communication with the team (surgical dentist, restorative dentist, laboratory and patient) should avoid this problem entirely.  If under-engineering cannot be avoided, then reduce the occlusion as described above.

bottom of page